Prompt gamma activation
The facility for the measurements of 10B concentrations in biological samples includes HPGe detector with 25% relative efficiency and associated Pb - 6Li2CO3 shielding. Described facility is installed at the distance of 1 m from the exit of the neutron guide. At the target position the neutron flux is approximately 3×106 n cm-2 s-1 at the reactor power of 8 MW. This facility makes it possible to determine 10B concentration of 1 ppm in 1 ml samples with statistical uncertainty of 5% within 15 min. This instrument is in a common property with Nuclear Research Institute, plc.
Applications:
The experimental set-up was mainly developed for the on-line determination of 10B concentrations of ppm-order in biological samples. This facility can also be used for determination of other isotopes with a sufficiently large (n,g) cross-section.
 
 Instrument  Parameters 
 
| Neutron flux | 3·106 n/cm-2 s-1 | 
| Beam cross section at target position | 25 x 5 mm2 | 
| Detector | HPGe (25% relative efficiency) | 



 
			 
                 
				
								
 
										 
		