CAS facebook.png twitter.png linkedIn.png
     

 



Head of laboratory
doc. Ing. Michal Vopálenský, Ph.D.
+420 567 225 343
 
Research Equipment
 

RTG1.png

Advanced workstation computer X-ray tomography combines two pairs of " X-ray tube - Display " (the Dual Source CT - DSCT ) in an orthogonal arrangement , which has a two-fold acceleration of the process of collecting data for tomographic reconstruction . The workplace has a fully motorized axes for distance setting "RTG - sample - detector " . This makes it possible to change the magnification of about 1.2 times to 100 times . Given the size of the detector pixels can change the resolution CT reconstruction from 0.2 millimeters to micrometers , the size of the detector. Very stable high resolution is possible also with regard to the use of anti-vibration table, on which the whole assembly is placed , and last but not least, thanks to the installation of high precision rotary table tomography .

RTG2.png

Another advantage is the ability to work DSCT parallel imaging an object in two spectra of X-ray radiation ( so-called dual energy radiography ) . This procedure allows you to highlight differences between the material components to the full X-ray spectrum similar to the attenuation of the radiation . If the sample consists of only two materials , these materials can be clearly distinguished. For multicomponent materials can only emphasize the differences .
 
Departments will also be equipped with detectors type Medipix / Timepix that allow you to set the display range in 4 -bit range. This will be the combination of two successive detectors collecting data 2x16 to get energy channels and tomographic reconstructions clearly distinguish up to 32 different materials.
 
This work is in DSCT completely unique , even in the world .
 
The main components of hardware and instrumentation:
X-ray tube XWT 160 TCHR :                                                                          X-ray tube XWT 240 SE:
· The micro- CT measurements                                                                   · For high power
· The minimum spot size of 1 micron                                                          · Minimum 4 micron spot size
· Maximum power on the target of 3 W                                                        · Maximum power 280 W target
· The maximum energy 160 keV                                                                   · The maximum energy of 240 keV

X-ray tube XWT 240 SE is designed for high- CT measurement and is therefore suitable for large samples or samples that strongly absorb X-rays . The diameter of the radiation beam based on the X-ray tube , the spot size is 4 µm. X-ray tube produces radiation with a maximum energy of 240 keV and its maximum power is 280 W.
X-ray tube XWT 160 TCHR is designed for micro- CT measurement and is therefore more suitable for smaller samples and those that do not absorb X-rays too strongly , because it produces less radiation of maximum energy and 160 keV. This X-ray tube operates in two modes depending on the desired application . High resolution mode provides a spot size of 1 µm , but the maximum power on the target of only 3 watts High energy mode provides a spot size of 4 µm , but the maximum power is 10 W.
 
Flat panel detectors                                                                                  WidePix detektor
· Size 400 x 400 mm                                                                                   · Size : 140 x 140 mm
· The matrix of pixels: 2048 x 2048                                                          · Matrix of pixels: 2560 x 2560
· Size of pixel: 200 x 200 µm                                                                      · Size : 55 x 55 micron
· Indirect signal conversion                                                                       · unique in the world
 

RTG5a6.jpg

RTG7.png

ABRT rotary table - 150 -AS :
· Resolution 0.55 arcseconds
· Maximum speed of 1200 rpm
· The weight capacity of 20 kg axially
· The weight capacity of 3 kg radially
 
Rotary table APR - 150 - DR - 135 :
· Resolution 0.08 arcseconds
· Maximum speed 600 rpm
· The weight capacity of 45 kg axially
· The weight capacity of 32 kg radially